Isolatorer
We offer isolators/glove boxes for sterility testing and product protection under aceptic conditions, BSL-3 work, and weighing and handling of hazardous materials.
Insulator for weighing
These isolators are suitable for work with API (active pharmaceutical ingredient)/ HAPI (hazardous active pharmaceutical ingredient), e.g. weighing, sampling and handling of hazardous material, where personal protection is necessary.
- Dual-chamber weighing isolator
- Airlock with one glove allowing for loading of material and instruments
- Working chamber with three gloves intended for work with hazardous materials
- Possible to ensure inert atmosphere inside chambers (N2, Ar)
- Leak tightness class 3 according to ISO 10648-2
- “C” Class cleanliness according to EU GMP standard
- “Audit Trail” compliant
- Electronic control system for automatic adjustment of basic operating modes managed by Siemens PLC
- Color touchscreen controls
- Jacket material: stainless steel AISI 304
- Isolator working chamber material – AISI 316L with a thickness of 3.00 mm
- Polished surface finish, Ra < 0.6 µm
- Inlet filtration HEPA H14 – “C” Class cleanliness
- Safe filtration of exhaust air with double HEPA H14 filters
- Easy-to-clean inner and outer surfaces
- Sliding trays between chambers for easy material transfer
- Oval holes with gloves for improved material handling
- Turbulent flow
- Negative pressure mode
Turbulent flow
Negative pressure mode
Aseptic isolator
These isolators are suitable for work with sterility testing and aseptic work where product protection is necessary.
- Single-chamber isolator for sterility testing
- Chamber with 4 sleeves for work with sterile materials
- Possible to add airlock chamber and provide sterilization with with H2O2
- Leak tightness class 3 according to ISO 10648-2
- “A” class cleanliness according to EU GMP
- “Audit Trail” Compliant
- Option to decontaminate with H2O2
- Electronic control system for automatic adjustment of basic operating modes managed by Siemens PLC
- Colour touchscreen control
- Jacket material: stainless steel AISI 304
- Chamber material – AISI 316L with a thickness of 3.00 mm
- Polished surface finish, Ra < 0.6 µm
- Inlet filtration HEPA H14 – “A” Class cleanlinnes
- Easy-to-clean inner and outer surfaces
- Sliding trays in Airlock for easy material transfer (applicable with airlock)
- Oval holes with gloves for improved material handling
- Laminar flow 0,45 m/s ±20%
- Positive pressure mode
BSL-3 isolator
These isolatorers are suitable for BSL-3 work, when you need both personal and product protection, and weighing of material.
- Working chamber and a pair of material airlocks
- Options for two or four gloves in the working chamber
- Leak tightness class 3 according to ISO 10648-2
- “A” Class cleanliness according to EU GMP standard
- Option to decontaminate with H2O2
- Electronic control system for automatic adjustment of basic operating modes managed by a Siemens PLC
- HMI Colour touchscreen
- Progress visualisation of each of the production steps and current parameters
- Jacket material: S235JR steel grade, powder coated surface
- Chamber interior material: AISI 304 stainless steel, powder coated surface
- Work surface material: AISI 316L stainless steel, polished
- Inlet filtration HEPA H14 – “A” Class cleanliness
- Safe filtration of exhaust air with double HEPA H14 filters
- Easy-to-clean interior and exterior surfaces
- Sliding trays between Airlocks and the Main chamber for easy material transfer
- Oval flanges with gloves for improved material handling
- Compact dimensions – easy relocation through a doorway in one piece
- Laminar flow 0,45 m/s ±20%
- Negative pressure mode
Laminar flow
Negative pressure mode